Electromagnetic compatibility
Guidelines and manufacturer’s declaration – electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or the
This device is intended for use in the electromagnetic environment specified below. The customer or the
user of this device should assure that it is used in such an environment.
Immunity test
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
discharge
± 8 kV air
discharge
Identical to
test level
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
Identical to
test level
Mains power quality should be that of a typical
commercial or hospital environment.
Surges
according to
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Identical to
test level
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
according to
IEC 61000-4-11
< 5% U
T
* (> 95%
dip in U
T
) for
0.5 cycle
40% U
T
(60%
dip in U
T
) for
5 cycles
70% U
T
(30%
dip in U
T
) for
25 cycles
< 5% U
T
(> 95%
dip in U
T
) for 5 s
Identical to
test level
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
this device requires continuous operation during
power mains interruptions, it is recommended
that the device is powered from an uninterruptible
power supply or a battery.
Magnetic
fields of mains
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
Identical to
test level
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
*Note: U
T
is the a.c. mains voltage prior to application of the test level.
DRAFT
–