5
Warnings (cont’d)
 
• The Zio AT system is MR Unsafe!
 
  - Do not expose the Zio AT patch or gateway to a magnetic 
resonance (MR) environment.
 
  - The Zio AT patch or gateway may present a risk of 
projectile injury due to the presence of ferromagnetic 
materials that can be attracted by the MR magnet core.
 
  - Thermal injury and burns may occur due to the metal 
components of the Zio AT patch that can heat during MR 
scanning.
 
  - The Zio AT patch may generate artifacts in the MR image.
 
  - The Zio AT patch or gateway may not function properly 
due to the strong magnetic and radiofrequency fields 
generated by the MR scanner.
If skin irritation such as severe redness, itching or 
allergic symptoms develop, remove the Zio AT Patch 
from the patient’s chest.  Call iRhythm Customer 
Service at 1.888.693.2401
CAUTION: Federal (USA) law restricts the sale of this 
device to or on the order of a physician.
 
Precautions
 
•  Safety and effectiveness of the Zio AT Patch on 
patients receiving any form of pacing therapy has not 
been established. Paced cardiac rhythms may not be 
accurately detected and may be incorrectly classified.
 
•  Safety and effectiveness of the Zio AT system on pediatric 
patients (younger than 18 years old) has not been 
established.
 
•  The Zio AT system includes temperature and humidity 
limitations when stored/transported. If exposed during 
storage/transport, patients may experience degradation 
of adhesive performance causing the Zio AT patch to slip 
or fall off during the patient wear duration.
 
•  The Zio AT system has a shelf-life date. Use of expired 
device may cause a degradation of ECG signal quality 
and/or low battery condition.
 
•  Do not use the Zio AT system if package is damaged. 
Device may not perform as intended.
 
•  Keep device and packaging away from young children.  
Contents may be harmful if swallowed.  Patch contains 
button cell batteries that are not accessible during 
normal use but, if exposed, are known choking hazards 
and may cause severe tissue injury if ingested.
 
•  Registration errors may result in limited functionality 
or erroneous ECG reporting.  Utmost caution should be 
applied to ensure that patient registration is accurate and 
complete.
The patient is an intended operator
Package Contents
1 Zio AT patch
1 Zio AT gateway, containing: 
 
1 postage-paid return envelope
1 Skin Prep & Placement Kit containing: 
 
1 patch card template 
 
1 disposable razor 
 
1 abrader disc 
 
4 alcohol wipes
1 Application instructions
1 Wearing your Zio manual & button press log containing: 
 
1 adhesive remover wipe