Orthofix Inc OFIX-5000-001 Manual De Usuario

Descargar
Página de 24
6
Visual and Audio Indicators
The LCD and audible beeps are designed to provide helpful information  
to the user. The screens, symbols, and beeps are explained below.
Treatment Screen
1:59
Treatment Screen – displays the treatment time remaining  
in hours and minutes. The timer counts down to zero until 
daily treatment is complete.
Exception  Screen
Device Expired
800-535-4492
E123
Exception Codes – Display of ERROR, any E codes  
(e.g., E01, E02), along with three slow beeps. Contact 
Patient Services at 800-535-4492 or 214-937-2718.
Charging Complete
Charging Complete – Indicates when the battery is 
fully charged.
Battery must be charged 
to turn on
E12345678
E12345678
Battery Empty – Indicates that the battery must be 
charged before treatment may continue.
Device Expired – Display of a closed lock indicates the 
device has been available for treatment for 365 days 
and will no longer provide treatment.
Compliance Screen
170/185 = 91.9%
Compliance Screen – Displays a compliance percentage 
which is calculated by the number of full treatments days 
completed over the number of available treatment days. 
The treatments days available begin once the device has 
been delivered to the patient and a minute of treatment 
time has been established.
Treatment Complete
Charging Screen
Battery Charging – Battery symbol filling repeatedly 
verifies that the device is charging.
Daily Prescribed Treatment complete 
1:59
u
u
u
u
u
u
u
u
u
Low Battery – Displays along with three fast beeps 
when recharging is recommended.
Low Battery Warning Screen