Bosch MIYN46 Manual De Usuario
14
en | Installation
MIYN46 Wall Transmitter
970.035b | F.01U.278.521 | v2.0 | 2012.12
User Manual US
Bosch Security Systems
4.2.8
Configuration and Wiring
Fig. 4.2
Configuration and Wiring
J4
J5
J6
1 2 3 4 5 6
1
2
3
4
5
Terminal J5
1 External 12-24 VDC power supply
1 External 12-24 VDC power supply
5-6 REL 2 Staff Presence
2 GND (0 V)
3-4 REL 1 Call for Help
3-4 REL 1 Call for Help
Terminal J6:
4-5 Reserve Call or Call for Assistance
3-4 Technical Alarm
1-2 Call for Help
Connector J4:
External Call for Help
trigger (pear push button)
External Call for Help
trigger (pear push button)
Solder Bridges (if soldered):
B8 Call for Assistance only by
B7
B6 No
B8 Call for Assistance only by
B7
B6 No
processing
B5 No Daily Message
B4 Not used
B3 Not used
B4 Not used
B3 Not used
Staff Presence
Call for Assistance from terminals J6[4-5]
Staff Presence
B1 REL 1 Call for Help
B2
B2
REL 2 Staff Presence
(o)
(o)
polarity
polarity
N
NC
N
NC
NC
NO
NC
NO
J2 Polarity of Technical Alarm input
J3 Polarity of
J3 Polarity of
input
terminal J6[4-5]
Connector J7:
Presence / Staff identification (optional)
Presence / Staff identification (optional)
N
NC
J1 External Call for Help trigger
input polarity
CR 2450
battery
Reset
NC-B2-NO NC-B1-NO
Solder bridges B1 and B2 are located on the bottom side of the PCB:
(o)
i
NOTICE!
Setting up one or more inputs to "NC" position results to higher power consumption. This con-
figuration should be used with external power supply or will reduce the battery life.