Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manual De Usuario

Los códigos de productos
MSO1074Z-S
Descargar
Página de 172
RIGOL 
2-122 
  DS1000Z Programming Guide 
:TRIGger:RS232:WIDTh 
Syntax 
:TRIGger:RS232:WIDTh <width> 
:TRIGger:RS232:WIDTh? 
Description 
Set or query the data bits when the trigger condition is DATA in RS232 trigger. 
 
Parameter 
Name 
Type 
Range 
Default 
<width> 
Discrete  {5|6|7|8} 
 
Return Format 
The query returns 5, 6, 7 or 8.   
Example 
:TRIGger:RS232:WIDTh 6 
Related 
Command
 
 
 
:TRIGger:RS232:BAUD 
Syntax 
:TRIGger:RS232:BAUD <baud_rate> 
:TRIGger:RS232:BAUD? 
Description 
Set or query the baud rate in RS232 trigger. The default unit is bps.   
Parameter 
Name 
Type 
Range 
Default 
<baud_rate>  Discrete 
{2400|4800|9600|19200|38400|57600|11
5200|230400|460800|921600|1000000|U
SER} 
2400 
 
Return Format 
The query returns an integer.   
Example 
:TRIGger:RS232:BAUD 4800 
Related 
Command
 
 
 
:TRIGger:RS232:BUSer 
Syntax 
:TRIGger:RS232:BUSer <user baud> 
:TRIGger:RS232:BUSer? 
Description 
Set or query the user-defined baud rate in RS232 trigger. The default unit is bps.   
Parameter 
Name 
Type 
Range 
Default 
<user baud>  Integer 
1 to 900000 
9600 
 
Return Format 
The query returns an integer.   
Example 
:TRIGger:RS232:BUSer 50000 
Related 
Command