:MEASure:ITEM 
Syntax 
:MEASure:ITEM <item>[,<src>[,<src>]] 
:MEASure:ITEM? <item>[,<src>[,<src>]] 
Description 
Measure any waveform parameter of the specified source, or query the measurement 
result of any waveform parameter of the specified source. 
Parameter
 
 
Name 
Type 
Range 
Default 
<item> 
Discrete 
{VMAX|VMIN|VPP|VTOP|VBASe|VAMP|VAVG| 
VRMS|OVERshoot|PREShoot|MARea|MPARea| 
PERiod|FREQuency|RTIMe|FTIMe|PWIDth| 
NWIDth|PDUTy|NDUTy|RDELay|FDELay| 
RPHase|FPHase|} 
——
 
<src> 
Discrete 
{CHANnel1|CHANnel2|CHANnel3|CHANnel4} 
——
 
 
Explanation 
  The [,<src>[,<src>]] parameter sets the source of the parameter to be measured.   
  If the parameter to be measured needs single source (such as VMAX, VMIN, VPP, 
VTOP, VBASe, VAMP, VAVG, VRMS, OVERshoot, MARea, MPARea, PREShoot, PERiod, 
FREQuency, RTIMe, FTIMe, PWIDth, NWIDth, PDUTy, NDUTy, etc.), one source 
requires to be set. If the [,<src>[,<src>]] parameter is omitted, the default source is 
selected by th
 command. 
 
  If the parameter to be measured needs two sources (such as RDELay, FDELay, 
RPHase, FPHase, etc.), two sources need to be sent with the command. If else, the 
command will be invalide. If the [,<src>[,<src>]] parameter is omitted, the default 
sources are selected by 
 and 
. 
 
Return 
Format 
The query returns the measurement result in scientific notation. 
Example 
:MEASure:ITEM OVERshoot,CHANnel2   /*Enable the overshoot measurement of CH2*/ 
:MEASure:ITEM? OVERshoot,CHANnel2   /*The query returns 8.888889e-03*/ 
Related 
Command