Rigol MSO1074Z-S 4-channel oscilloscope, Digital Storage oscilloscope, MSO1074Z-S Manual De Usuario
Los códigos de productos
MSO1074Z-S
RIGOL
2-38
DS1000Z Programming Guide
:MATH:OFFSet
Syntax
:MATH:OFFSet <offs>
:MATH:OFFSet?
:MATH:OFFSet?
Description
Set or query the vertical offset of the operation result. The unit dependents on the
operator currently selected and the unit selected by the source.
Parameter
Name
Type
Range
Default
<offs>
Real
Be related to the vertical scale of the operation result
Range: -1000*MathVerticalScale ~
1000*MathVerticalScale
Step: MathVerticalScale/50
0.00V
Explanation
MathVerticalScale is the vertical scale of the operation result and can be set by
the
Return
Format
The query returns the vertical offset of the operation result in scientific notation.
Example
:MATH:OFFSet 8 /*Set the vertical offset to 8V*/
:MATH:OFFSet? /*The query returns 8.000000e+00*/
Related
Command
:MATH:INVert
Syntax
:MATH:INVert <bool>
:MATH:INVert?
:MATH:INVert?
Description
Enable or disable the inverted display mode of the operation result, or query the inverted
display mode status of the operation result.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation
This command is invalid for FFT operation.
Return
Format
The query returns 0 or 1.
Example
:MATH:INVert ON
Related
Command
:MATH:RESet
Syntax
:MATH:RESet
Description
Sending this command, the instrument adjusts the vertical scale of the operation result to
the best value according to the current operator, the horizontal timebase of the source.
Related
Command