DS2000A User’s Guide
5-32
Setup Type:
Press SetupType to select the desired setup type.
Setup: set the time that the data stays stable and constant before the clock
Press SetupType to select the desired setup type.
Setup: set the time that the data stays stable and constant before the clock
edge appears. Press Setup to set the setup time and the range is from 2 ns to 1
s.
s.
Hold: set the time that the data stays stable and constant after the clock edge
appears. Press Hold to set the hold time and the range is from 2 ns to 1 s.
SetupHold: set the time that the data stays stable and constant before and
after the clock edge appears. Press Setup and Hold to set the setup time and
hold time respectively and the range is from 2 ns to 1 s.
hold time respectively and the range is from 2 ns to 1 s.
Trigger Mode:
) under this trigger type as auto,
normal or single. The corresponding status light of the current trigger mode turns on.
Trigger Setting:
Press Setting to set the trigger parameters (trigger holdoff and noise rejection)
under this trigger type.
Trigger Level:
Use TRIGGER
Trigger Level:
Use TRIGGER
LEVEL to modify the level. For details, please refer to the