DS2000A User’s Guide
3-6
4. As shown in the figure above, the distances from the crossing points of axis and
the circle to the origin of the coordinates are approximately equal. Thus, the
phase deviation angle θ=±arcsin1=90°.
phase deviation angle θ=±arcsin1=90°.
Note:
In Y-T mode, the oscilloscope could use any sample rate (within the guaranteed
range) to capture waveform. The maximum sample rate of X-Y mode is 1.0
GSa/s. Generally, reducing the sample rate properly could improve the display
effect of Lissajous figure.
GSa/s. Generally, reducing the sample rate properly could improve the display
effect of Lissajous figure.
The following functions are not available in X-Y mode:
Auto measure, cursor measure, math operation, reference waveform, delayed
sweep, vector display, HORIZONTAL POSITION, trigger control, memory
depth, acquisition mode, Pass/Fail test and waveform record.
sweep, vector display, HORIZONTAL POSITION, trigger control, memory
depth, acquisition mode, Pass/Fail test and waveform record.
Method 2: Use the shortcut measurement function
Please refer to “Phase A→B ” and “Phase A→B ” measurement functions of “Delay
and Phase” in page