Utility Menu - Page 2/4 
 
D
S
ELF 
C
AL
 
1.  Press the Do Self Cal option button to perform a self-calibration. 
The Do Self Cal screen is shown. 
 
2.  As the screen indicates, disconnect anything plugged into input 
connectors (except for the power cable) before pressing the 
SINGLE Front Panel (on page 10) button and performing a self-
calibration. Press the RUN/STOP front panel button to exit the Do 
Self Cal
 screen. 
3.  After pressing the SINGLE front panel button, a status bar shows 
the self-calibration progress for each channel on the instrument.