Voltcraft VC880 Digital Multimeter with Software included 40 000 counts CAT IV 600V, CAT III 1000V VC880 Hoja De Datos
Los códigos de productos
VC880
TECHNICAL DATA
Symbol
Max. 40000 counts (characters)
Measuring rate
approx. 2-3 measurements/second, Bargraf approx.
10 measurements/second
10 measurements/second
Measuring line length
approx. 90 cm each
Measuring impedance
>10M
Ω (V range)
Operating voltage
9V block battery
Working conditions
0 to 30°C (<75%rF), >30 to 40°C (<50%rF)
Operating altitude
max. 2,000 m
Storage temperature
-10°C to +50°C
Weight
approx. 380 g
Dimensions (LxWxH)
185 x 91 x 43 (mm)
Over-voltage category
CAT III 1,000 V, CAT IV 600 V, contamination degree 2
Measurement tolerances
Statement of accuracy in
Statement of accuracy in
± (% of reading + display error in counts (= number of smallest points)). The accuracy is valid
for one year at a temperature of +23°C
± 5°C, and at a relative humidity of less than 75 %, non-condensing.
Temperature coefficient: +0.1 x (specified accuracy)/1°C
Direct voltage
Range
Accuracy
Resolution
400 mV
±(0,03% + 10)
0,01 mV
4 V
±(0,05% + 10)
0,0001 V
40 V
0,001 V
400 V
0,01 V
1000 V
0,1 V
Overload protection 1000 V; Impedance: 10 M
Ω
Alternating voltage
Range
Accuracy
Resolution
Frequenz range
4 V
0,0001 V
±(0,5 % + 40)
45 Hz - 1kHz
±(1,2 % + 40)
1 kHz - 10 kHz
±(3 % + 40)
10 kHz - 20 kHz
±(4 % + 40)
20 kHz - 100 kHz
40 V
0,001 V
±(0,5 % + 40)
45 Hz - 1kHz
±(1,2 % + 40)
1 kHz - 10 kHz
±(3 % + 40)
10 kHz - 20 kHz
±(6 % + 40)
20 kHz - 100 kHz
400 V
0,01 V
±(0,5 % + 40)
45 Hz - 1kHz
±(1,2 % + 40)
1 kHz - 10 kHz
±(3 % + 40)
10 kHz - 20 kHz
Not specified
20 kHz - 100 kHz
1000 V
0,1 V
±(1,2 % + 40)
45 Hz - 1kHz
±(3 % + 40)
1 kHz - 5 kHz
±(6 % + 40)
5 kHz - 10 kHz
Not specified
10 kHz - 100 kHz
Overload protection 1000 V; Impedance: 10 M
Ω
TrueRMS in the measuring range of 10 – 100%: Crest factor: max. 3.0 (at 750V max. 1.5)