Chauvin Arnoux P01157020 Simple Logger P01157020 Manual De Usuario
Los códigos de productos
P01157020
SIMPLE LOGGER II
English
Page 43
MODEL
L111
Electrical
Input
One
Input connection
Two 4mm safety banana jacks
Input level
0 to 1A
AC
Uncertainties (50 / 60 Hz)
Resolution
0 to 10 mA: unspecified
10 to 50 mA: ± (0,5% of reading + 1 mA)
50 to 1000 mA: ± (0,5% of reading + 0,5 mA)
0.1 mA
Maximum input level
1,2 A
Input impedance
1
Ω
Sample rate
64 samples/cycle
Storage rate
Programmable from 125 ms to 1 day
Storage modes
Start/Stop, FIFO or Extended Recording Mode* (XRM™)
Recording length
Programmable using DataView
®
from 15 minutes to 8 weeks
Memory
240,000 measurements (512 kB). The recorded data is stored in
non-volatile memory retained even if the battery is low or removed.
Communication
USB 2.0 optically isolated
Power source **
2 x 1.5 V AA (LR6) cell Alkaline batteries
Battery life
100 hours to >45 days (dependent on storage rate/recording length)
Mechanical
Dimensions
136x70x32 mm (5.38x2.75x1.28")
Weight (with batteries)
180 g (6.4 oz)
Case
UL94-V0
Vibration
IEC 68-2-6 (1.5 mm, 10 to 55 Hz)
Shock
IEC 68-2-27 (30 G)
Drop
IEC 68-2-32 (1 m)
Environmental
Operating temperature
- 10°C to + 50°C (14°F to 122°F)
Storage temperature
- 20°C to + 60°C (- 4°F to + 140°F)
Relative humidity
up to 85% at 35°C (95°F) without condensation
Altitude
2000 m
Safety
Conformity
EN 61010-1; 50 V CAT III; Pollution level 2
Level of protection
IP 40
*See the end of chapter