APPENDIX 
71 
Output Limitation for 
Withstanding Voltage Testing 
Upper Current 
Pause 
Output Time 
15mA≤I≤20mA 
At least as long 
as the output 
time 
Maximum 180 
seconds 
0.01mA≤I<15mA  Not necessary  Continuous 
output 
possible 
 
DC Withstanding Voltage 
Output Voltage Range 
0.10kV~ 6.00kV 
Output Voltage Resolution 
10V 
Maximum Rated Load 
25W (5kV/5mA)  
Maximum Rated Current 
6mA 
0.01mA ~ 2mA (0.1kV≤V≤0.5kV) 
0.01mA ~ 6mA (0.5kV<V≤6kV) 
Voltmeter Accuracy 
±(1.5% of reading + 2 counts) 
Output Voltage Accuracy 
±(1.5% of setting + 2 counts) with no load 
Voltage Regulation 
± 1.5% +2 counts 
[Maximum rated load → no load] 
Current Measurement Range  0.01mA~6.00mA  
Current Best Resolution 
10uA 
0.01mA(0.01mA~6.00mA)  
Current Measurement 
Accuracy 
± 2% of reading + 5 counts when I <1.00mA  
± 2% of reading + 3 counts when I ≥1.00mA 
Current Judgment Accuracy 
± 3% of setting + 5 counts when I <1.00mA  
± 3% of setting + 3 counts when I ≥1.00mA  
Window Comparator Method  Yes 
ARC DETECT 
Yes 
RAMP (Ramp Time) 
0.1S fixed 
TIMER (Test Time) 
OFF, 1S~180S 
GND 
ON 
Judgment Valid Range 
H/L Setting: 0.10mA to 6.00mA 
 
Insulation Resistance Test 
Output Voltage 
50V, 100V, 250V, 500V, 1000V 
Output Voltage Accuracy 
±(3% of setting+1 count) with no load 
Resistance Measurement 
Range 
1M
 ~ 2000M
 
Test Voltage  
Measurement Range  Accuracy 
 
50V/100V/250 
1 ~ 50MΩ 
±(5% of reading + 2MΩ) 
51~2000MΩ
 
±(10% of reading + 2MΩ) 
 
500V/1000V 
1 ~ 500MΩ 
±(5% of reading + 2MΩ) 
501~2000MΩ
 
±(10% of reading + 2MΩ) 
Voltmeter accuracy 
±(3% of reading + 1 count) 
Continued…