Gw Instek GPT-9602 Insulation measuring device, GPT-9602 Manual De Usuario
Los códigos de productos
GPT-9602
GPT-9600 Series User Manual
44
Example
H = 2 2 . 0 m A
m A
L
0
=
0 0
.
0 8 S
= 0
T
CW
D
A CW
R
I
D
O E
M
T M E
I
TEST Status
Measured current
TEST voltage
Remaining test time
0 94
k V
S
E
T
T
A
m
9 00
Stop the Test
9. To stop the test at any time when it
is running, press the STOP button.
The test will stop immediately.
When the STOP button is pressed,
a judgment is not made on the test.
The test will stop immediately.
When the STOP button is pressed,
a judgment is not made on the test.
All panel keys except the STOP
button are locked when the tester
is in STOP status.
button are locked when the tester
is in STOP status.
STOP
Example
H = 2 2 . 0 m A
m A
L
0
=
0 0
.
0 8 S
= 0
T
CW
D
A CW
R
I
D
O E
M
T M E
I
STOP Status
0 94
k V
O
T
S
P
A
m
9 00
10. To put the tester back into READY
status, press the STOP button
again.
again.
STOP
Note
Do not touch any terminals, test leads or any other
connections when the test is on.
connections when the test is on.