F
AQ
• The tester will not turn on.
• The panel keys are not working.
• When I press the START button the tester will not start testing?
• The accuracy does not match the specification.
• The panel keys are not working.
• When I press the START button the tester will not start testing?
• The accuracy does not match the specification.
The tester will not turn on.
Ensure the power cord is connected. Ensure the line input is set to
the correct line voltage. Check to make sure the fuse is not blown.
See page 146.
the correct line voltage. Check to make sure the fuse is not blown.
See page 146.
The panel keys are not working.
Ensure the tester is not in remote mode, page 107.
Ensure the tester is not in SIGNAL I/O or Remote Connect mode,
page 76.
Ensure the tester is not in SIGNAL I/O or Remote Connect mode,
page 76.
When I press the START button the tester will not start
testing?
testing?
The tester must first be in the READY status before a test can be
started. Ensure the tester displays READY before pressing the
START button, page 58 (manual test), 88(automatic test).
If “Double Action” is enabled, the START button must be pressed
0.5 seconds after the STOP button is pressed, otherwise the tester
will not start testing.
started. Ensure the tester displays READY before pressing the
START button, page 58 (manual test), 88(automatic test).
If “Double Action” is enabled, the START button must be pressed
0.5 seconds after the STOP button is pressed, otherwise the tester