GPT-9000 Series User Manual 
44
 
 
I R  
G B  
 
m A
A CW
D CW
I / L
H
O
I M E
T
R
 
 
E        
 
 
F R E Q =  
0 H z  
6
0
   
100
k
 
V
 
       
R
A C
 
=
F F
O
I
D
E
 
 
                         
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
M
M A U
N
N A
_
1 . 0 0 m A
0
L
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
 
 
T
2
M A N U = * * * - 0 0
cursor  
 
2.  Use the scroll wheel to set the 
TIMER value. 
 
 
 
ACW 
DCW 
IR  
GB 
000.5s~999.9s  
000.5s~999.9s 
001.0s~999.9s 
000.5s~999.9s 
 Note 
With the ACW test function, when the test current 
is between 30mA and 40mA, the ramp time + test 
time cannot exceed 240 seconds. At this current 
level, the tester also needs to pause after a test for 
a time equal to or greater than the output time. See 
the specifications on page 149 for details. 
Special Manual 
Mode 
When in special MANU test mode (page 67) the 
Timer can be turned off when using the DCW 
or ACW test function. 
Hold the TIMER soft-key for 3 
seconds to turn the timer off. 
I M E
T
R
 
 Note 
The timer can only be turned off under special 
MANU test mode, however there is a limitation:  
The timer cannot be turned off (limited to 240s) if 
the test current is between 30mA and 40mA in 
ACW mode.  
The discharge time and initial test time cannot be 
edited.