Example
(in AUTO mode)
(in AUTO mode)
MEAS10?
>IR, FAIL ,0.225kV ,999M ohm,T=010.3S
Returns step 10 of the current automatic result.
>IR, FAIL ,0.225kV ,999M ohm,T=010.3S
Returns step 10 of the current automatic result.
MAIN:FUNCtion
Set
Query
Description
Changes the mode between AUTO and MANU.
Syntax
Query Syntax
Query Syntax
MAIN:FUNCtion {MANU|AUTO}
MAIN:FUNCtion ?
MAIN:FUNCtion ?
Parameter/
Return parameter
Return parameter
MANU
Puts the tester mode to MANU.
AUTO
Puts the tester mode to AUTO.
Example
MAIN:FUNC MANU
Sets the tester to MANU mode.