GPT-9000 Series User Manual
50
3. Use the scroll wheel to set PASS
HOLD.
PASS HOLD
OFF, ON
4. Press the EDIT/SAVE key to save
and exit the MANU Utility menu.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the
MANU Utility menu to cancel and exit.
MANU Utility menu to cancel and exit.
Setting FAIL HOLD
Background
The FAIL HOLD settings only apply to the
selected test in AUTO tests. When the FAIL
HOLD setting is set to ON, a FAIL judgment is
held until the START button is pressed.
selected test in AUTO tests. When the FAIL
HOLD setting is set to ON, a FAIL judgment is
held until the START button is pressed.
Note
The FAIL HOLD setting only applies to AUTO
tests. This setting is ignored when running a
single MANU test.
tests. This setting is ignored when running a
single MANU test.
Steps
1. Press the UTILITY key on the front
panel when the tester is in EDIT
status. The display will go from the
normal EDIT status to the MANU
Utility menu for the current test.
status. The display will go from the
normal EDIT status to the MANU
Utility menu for the current test.