Greisinger GDUSB 1000 Universal USB Interface Adapter for GMSD Pressure Sensors 601666 Manual De Usuario

Los códigos de productos
601666
Descargar
Página de 14
S56.0.01.6C-04 
 
page 4 of 7 
 
 
 
 
 
5 Operating modes 
After restart the adapter is in the standard-mode. 
5.1  Standard - Mode 
In this mode the GDUSB1000 acts like a portable measuring instrument of the GMH3xxx-series.  
It gives back the wanted measurement on request. 
For this purpose 4 measuring channels are allocated at the adresses 1 – 4: 
1   Measurand: 
with adjustable low-pass filter (0.00 ... 10.00) 
2   Average:   
the average between two measuring-points 
3   Minimal peak: 
lowest measurand for the last request; unfiltrated, with highest sample rate 
4   Maximal peak: 
highest measurand for the last request; unfiltrated, with highest sample rate 
 
The communication can occur with every standard software for the GMH3xxx or 
EASY
BUS
 (e.g. 
EBS9M) and offers the possibility to monitor long-run measurements of pressure for example. 
For the integration in your own programs, we offer you the function-library EASYBUS.DLL. 
This operating mode is especially adapted for the measuring of 1 measurand per second. 
For example: long-run measurements of pressure, leakage test,... 
Note: for use the GDUSB1000 you must select in the software the interface USB3100. 
5.2  Fast - Mode 
The device can be set to the mode “fast measurand display” via command. In this mode the measure-
ment is started by a trigger condition. Then the device automatically gives back the measurands in the 
adjusted cycle time without any request, until the stop condition is fulfilled. 
Adjustable parameters are among other ones: 
Cycle time:  
1ms ... 10sec.; programmable in steps of ms 
Measuring channel: 
cycle < 5 ms   it is possible to choose the recording channel (measurand, min. 
peak, max. peak) 
 cycle 
> 5 ms   all four measuring channels are recorded automatically 
Trigger conditions: 
5 for the measuring start   
 
 
 
6 for the measuring stop 
Pre-trigger: 
50 data sets (at a cycle time of 
≥ 5ms) respectively 200 measurands (cycle 
time < 5 ms) 
 
For the communication and configuration of the device in this mode the software GSOFT-USB is nec-
essary. 
This mode is especially adapted for measuring very fast pressure changes (1000 per second) 
For example: detecting pressure peaks at switching operations, ....