Gw Instek GPT-9904 Insulation measuring device, GPT-9904 Manual De Usuario

Los códigos de productos
GPT-9904
Descargar
Página de 171
 
FAQ 
155 
F
AQ 
 
•  The tester will not turn on. 
•  The panel keys are not working. 
•  When I press the START button the tester will not start testing? 
•  The accuracy does not match the specification. 
The tester will not turn on. 
 
Ensure the power cord is connected. Ensure the line input is set to 
the correct line voltage. Check to make sure the fuse is not blown. 
See page 157. 
The panel keys are not working. 
 
Ensure the tester is not in remote mode, page 114.  
Ensure the tester is not in SIGNAL I/O or Remote Connect mode, 
page 100.  
When I press the START button the tester will not start 
testing? 
 
The tester must first be in the READY status before a test can be 
started. Ensure the tester displays READY before pressing the 
START button, page 62 (manual test), 88(automatic test). 
If “Double Action” is enabled, the START button must be pressed 
0.5 seconds after the STOP button is pressed, otherwise the tester 
will not start testing.