OPERATION 
41 
Steps 
1.  Press the UP / DOWN arrow keys 
to bring the cursor to the voltage 
setting. 
 
 
I R  
G B  
 
A CW
D CW
I / L
H
O
I M E
T
R
 
 
E        
 
 
F R E Q =  
0 H z  
6
0
   
100
k
 
V
 
       
 
 
 
 
 
   
 
I
D
E
 
 
                         
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
M
M A U
N
N A
_
1 . 0 0 m A
I
H
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
 
 
T
2
M A N U = * * * - 0 0
cursor
m A
 
 
2.  Use the scroll wheel to set the 
voltage level.  
 
 
 
ACW 
DCW 
IR 
GB 
0.100kV ~ 5kV 
0.100kV ~ 6kV 
0.05kV ~ 1kV (50V steps) 
3.00A ~ 30.00A (GPT-98XX) 
3.00A~ 32.00A (GPT-99XX) 
 Note 
When setting the voltage, be aware that a 
maximum of 200VA can be set for ACW and 50W 
for DCW (GPT-98XX) or 500VA and 100W, 
respectively for GPT-9900. 
 
The ground bond voltage (GBV) is calculated as 
the HI SET limit x Test Current.  
Setting the Test Frequency 
 
Background 
A test frequency of 60Hz or 50Hz can be set, 
regardless of the input line voltage.  The test 
frequency setting only applies to ACW and GB 
tests.