GPT-9000 Series User Manual 
46
 
Setting the Test Time (Timer) 
 
Background 
The TIMER setting is used to set the test time 
for the current test. The test time determines 
how long the test voltage or current is applied 
to the DUT. This test time does not include 
Ramp , initial start time or discharge time 
(note: GB does not have Ramp  or discharge 
times). The test time can be set from 0.5 seconds 
to 999.9 seconds for ACW, DCW and GB and 
1.0 second to 999.9 seconds for IR, with a 
resolution of 0.1 seconds for all modes.  The 
timer can be turned off when in the special 
MANU test mode when using the ACW or 
DCW test functions.  
Each test has an initial test time of 100ms and a 
discharge time (except GB). The total discharge 
time depends on the DUT and test voltage. 
Discharge time 
Test V
time
Start V
Initial time 
(100ms)
TEST TIME
RAMP
 
Steps 
1.  Press the TIMER soft-key or use 
the UP/DOWN arrow keys to 
bring the cursor to the TIMER 
setting. 
I M E
T
R
 
OR