H57.0.0X.6C-01 
Operating Manual 
MSD … 
page 4 of 4
 
 
Specification 
 
6.1  Specification (
MSD … BAE): 
 
MSD 1 BAE 
MSD 2.5 BAE 
MSD 4 BAE 
MSD 6 BAE 
MSD … BAE 
(special range)
 
Measuring range: 
… 1000 mbar abs.  0 … 2500 mbar abs.  0 … 4000 mbar abs.  0 … 6000 mbar abs. 
refer to type plate
 
Overload:  (max.) 
5 bar abs. 
10 bar abs. 
17 bar abs. 
35 bar abs. 
refer to type plate
 
Resolution: 
1 mbar 
1 mbar 
1 mbar 
1 mbar 
refer to type plate
 
Sensortyp: 
stainless steel absolute pressure sensor.  
Suitable for aggressive media, water, etc. 
6.2  Specification (
MSD … ..RE): 
 
M
S
D
 400 
M
R
E
 
M
S
D
 1 
B
R
E
 
M
S
D
 2.
B
R
E
 
M
S
D
 4
 BRE
 
M
S
D
 6 
B
R
E
 
M
S
D
 10 
B
R
E
 
M
S
D
 25 
B
R
E
 
M
S
D
 40 
B
R
E
 
Measuring range: 
0,0 
… 400,0 
mbar rel. 
… 1000 
mbar rel. 
… 2500 
mbar rel. 
… 4000 
mbar rel. 
… 6000 
mbar rel. 
0,00 
… 10,00 
bar rel. 
0,00 
… 25,00 
bar rel. 
0,00 
… 40,00 
bar rel. 
Overload:  (max.) 
2 bar 
5 bar 
10 bar 
17 bar 
35 bar 
35 bar 
50 bar 
80 bar 
Resolution: 
0,1 mbar 
1 mbar 
1 mbar 
1 mbar 
1 mbar 
0,01 bar 
0,01 bar 
0,01 bar 
 
 
M
S
D
 60 
B
R
E
 
M
S
D
 100 
B
R
E
 
M
S
D
 160 
B
R
E
 
M
S
D
 250 
B
R
E
 
M
S
D
 400 
B
R
E
 
M
S
D
 600 
B
R
E
 
M
S
D
 1
0
00
 B
R
E
 
MSD … ..RE 
(Sonderbereich)
 
Measuring range: 
0,00 
… 60,00 
bar rel. 
0,0 
… 100,0 
bar rel. 
0,0 
… 160,0 
bar rel. 
0,0 
… 250,0 
bar rel. 
0,0 
… 400,0 
bar rel. 
0, 0 
… 600,0 
bar rel. 
… 1000  
bar rel. 
refer to type plate
 
Overload:  (max.) 
120 bar 
200 bar 
320 bar 
500 bar 
800 bar 
1200 bar 
1500 bar 
refer to type plate
 
Resolution: 
0,01 bar 
0,1 bar 
0,1 bar 
0,1 bar 
0,1 bar 
0,1 bar 
1 bar 
refer to type plate
 
Sensortyp: 
stainless steel relative pressure sensor for overpressure measuring. 
Suitable for aggressive media, water, etc. 
Caution: The pressure compensation hole at the back part of the housing has to be kept clean!  
6.3  Common specifications (
MSD … BAE, MSD … ..RE): 
Accuracy: 
(typ. values) 
±0,2%FS  (hysteresis and linearity)
 
±0,4%FS
 
 (temperature influence from 0-50°C) 
Pressure connection:  connections thread G1/4. Key width: 27 mm 
Device Connection: 
M12-plug,  for connection cable MSD-K31 
Electronics: 
PC-board with amplifier and data memory for sensor data  
(measuring data, calibration etc.) integrated in sensor housing. 
Nominal temperature:  25 °C 
 
Operating conditions:  -20 to +70 °C  (compensated range: 0 to 70 °C) 
Storage temperature: 
-40 to +80 °C 
Housing:   
made of stainless steel (CrNi steel or Elgiloy®) 
Weight: 
 
220 g 
IP rating:   
IP 67 (sensor), IP 54 (plug) 
EMC: 
The MSD ... corresponds to the essential protection ratings  
established in the Regulations of the Council for the Approximation  
of Legislation for the member countries regarding electromagnetic  
compatibility (2004/108/EG). Additional fault: <1%