8 Specification
Measuring ranges:
Display range:
Display range:
-10.0 to 350.0 mbar relative (under pressure down to -350.0 mbar, p.r.t. chapter 5)
Overload:
max. 1000 mbar rel. (without destruction or recalibration of sensor being necessary)
Resolution:
0.1 mbar
Pressure units:
mbar, bar, kPa, MPa, mmHg, PSI, mH
2
O
(display ‚m‘)
, selectable
Accuracy: (typ.)
±0,2%FS (hysteresis and linearity)
±0,4%FS (temperature dependency 0-50°C)
±0,4%FS (temperature dependency 0-50°C)
OPTION higher accuracy:
±0.1%FS (hysteresis and linearity)
±0.4%FS (temperature dependency 0-50°C)
±0.4%FS (temperature dependency 0-50°C)
Measuring rate:
slow:
4 meas./sec (ConF-Rate = Slow)
fast:
>1000 meas./sec (ConF-Rate = FASt and P.dEt)
Nominal temperature:
25°C
Sensor:
Piezo-resistive relative pressure sensor integrated in device.
Suitable for air and non-corrosive and non-ionizing gases and liquids.
(Not suitable for water – use air buffering)
Suitable for air and non-corrosive and non-ionizing gases and liquids.
(Not suitable for water – use air buffering)
Connection:
2 metal pressure ports for connection to 6 x 1 mm tubes at the top of device (4mm
inner tube Ø)
inner tube Ø)
Logger:
2 Functions: individual value logger („Func–Stor“) and cyclic logger („Func–CYCL“)
Memory:
Stor: 99 data sets
CYCL: 10000 data sets (in max. 64 recording sequences)
CYCL: 10000 data sets (in max. 64 recording sequences)
Cycle time CYCL:
1...3600 seconds
Display:
2 four digit LCDs (12.4mm high and 7 mm high) for measuring values, and for min/
max memories, hold function, etc. as well as additional functional arrows.
max memories, hold function, etc. as well as additional functional arrows.
Pushbuttons:
6 membrane keys
Output:
3.5 mm audio plug, stereo
Output function:
selectable as serial interface or analog output
Interface:
Serial interface (3.5mm jack) can be connected to RS232 or USB interface of a PC
via electrically isolated interface adapter GRS3100, GRS3105 or USB3100 (see
accessories).
via electrically isolated interface adapter GRS3100, GRS3105 or USB3100 (see
accessories).
Analog output:
0 ... 1 Volt, freely scaleable (resolution 12 bit)
Power supply:
9V battery, type: IEC 6F22 (included in scope of supply)
as well as additional d.c. connector (diameter of internal pin 1.9 mm) for external
10.5-12V direct voltage supply. (suitable power supply: GNG10/3000)
as well as additional d.c. connector (diameter of internal pin 1.9 mm) for external
10.5-12V direct voltage supply. (suitable power supply: GNG10/3000)
Power consumption:
Slow measuring rate: ~ 0.6 mA
Fast measuring rate: < 2.5 mA
Low-Power-Logger: < 0.1 mA (for cycle time>30s, without interface
communication active and no alarm horn sounding) up to 0.4 mA (at cycle time 1s)
Fast measuring rate: < 2.5 mA
Low-Power-Logger: < 0.1 mA (for cycle time>30s, without interface
communication active and no alarm horn sounding) up to 0.4 mA (at cycle time 1s)
Low battery warning:
' bAt '
Additional Functions:
Power-Off-Function:
Power-Off-Function:
Device will be automatically switched off if no key is pressed/no interface
communication takes place for the time of the power-off delay. The power-off delay
can be set to values between 1 and 120 min.; it can be completely deactivated.
communication takes place for the time of the power-off delay. The power-off delay
can be set to values between 1 and 120 min.; it can be completely deactivated.
Min/Max-Alarm:
The measuring value is constantly monitored for the min and max rails set.
Alarming is done by integrated horn, display and interface
Alarming is done by integrated horn, display and interface
Real time clock:
Integrated clock with date and year
_
+