Greisinger GMH3161-12 Digital Fine Manometer 605296 Manual De Usuario

Los códigos de productos
605296
Descargar
Página de 24
H61.0.2X.6C-01 
User’s Manual GMH 3161-… 
page 10 of 12 
 
 
 
 
11  Error And System Messages 
Display 
Meaning 
What to do? 
 
Low  battery power, device will only continue 
operation for a short period of time 
Replace battery 
 
Battery empty  
Replace battery 
Mains operation without battery: wrong voltage 
Check  power supply, replace it when 
necessary 
No display or 
confused 
characters, 
device does 
not react on 
keypress 
Battery empty 
Replace battery 
Mains operation without battery: wrong voltage 
or polarity 
Check  power supply, replace it when 
necessary 
System error 
Disconnect battery and power supplies, 
wait shortly, then reconnect 
Device defective 
Return to manufacturer for repair 
Err.1 
Measured value above allowable range 
Check: pressure above max. range?  
-> measuring value to high 
Sensor defective 
Return to manufacturer for repair 
Err.2 
Measured value below allowable range 
Check: pressure below min range?  
-> measuring value to low 
Sensor defective 
Return to manufacturer for repair 
Err.4 
Value is too low to be displayed, tara is set 
Check: display below -2000 (tara?)?  
Err.9 
Measured value far out of allowable range 
Check: pressure not within sensor range?  
Err.7 
System error 
Return to manufacturer for repair