18 Specifications
Display range
maximal -19999...19999 digit, depends on used sensor
Number of channels
1
Suitable probes
GMSD
… - K51 and MSD … sensors with connection cable MSD-K51
Available ranges/resolutions
ranging from -
1.999 … 2,500 mbar / 0.001 mbar to 0 … 1000 bar / 1 bar
Connections
sensor
7-pole bayonet connector
output /
ext. supply
ext. supply
4-pole connector for serial interface and supply (USB Adapter USB 5100)
analog output 0-1V, adjustable
analog output 0-1V, adjustable
Display units
depends on meas. range selection (depends on connected sensor):
mbar, bar, Pa, kPa, MPa, mmHg, inHg, PSI, mH
mbar, bar, Pa, kPa, MPa, mmHg, inHg, PSI, mH
2
O
User unit
freely adjustable
Measuring frequency
4 / s or 1000 / s
Accuracy
± 0.1 % FS ± 1 digit
Working conditions
-25 ... 50 °C; 0 ... 95 %RH (non condensing)
Storage conditions
-25 ... 70 °C
Display
4 ½ digit 7-segment display, illuminated (white)
Correction function
offset / slope (via menu),
if offset/slope value not equal zero: "corr" is displayed
if offset/slope value not equal zero: "corr" is displayed
Leakage test function
Pressure change rate display, configurable time base
Air velocity / volume flow
measuring with pitot tube
Real-time clock
integrated clock with date and year
Data logger
cyclic
10000 data sets
manual
1000 data sets (with measuring point input,
40 adjustable measuring point text numbers)
40 adjustable measuring point text numbers)
Alarm
channels
1 with individual alarm boundaries
types
buzzer/visual/interface
Additional functions
min- / max- / hold- function
Averaging
adjustable
, 1 … 120 seconds
Housing
protection class IP65 and IP67
dimensions
L*W*H [mm]
L*W*H [mm]
160 * 86 * 37 incl. protective covering,
approx. 250 g incl. batteries and protective covering
Power supply
2*AAA-batteries (included in delivery)
current
consumption
consumption
2.0 mA (at Out = Off, equivalent to 500 h),
illumination ~10mA (turns off automatically)
illumination ~10mA (turns off automatically)
Battery-change indicator
automatically if battery exhausted "bAt", warning "bAt" flashing
Auto-off-function
Device will be automatically switched off if no key is pressed/no interface
communication takes place for the time of the power-off delay.
The power-off delay can be set to values between 1and 120 min.; it can be
completely deactivated.
communication takes place for the time of the power-off delay.
The power-off delay can be set to values between 1and 120 min.; it can be
completely deactivated.
EMC
The device corresponds to the essential protection ratings established in the
Regulations of the Council for the Approximation of Legislation for the member
countries regarding electromagnetic compatibility (2004/108/EG)
Additional fault: <1%
Regulations of the Council for the Approximation of Legislation for the member
countries regarding electromagnetic compatibility (2004/108/EG)
Additional fault: <1%