II.) Average ambient absolute pressure „PAbS“ (function of the altitude above sea level): 
Necessary for a correct evaluation of the state of the electrode. If the application works for example at an altitude of 
350 m above sea level, 980mbar is a correct setting. Please refer to Appendix B 
4.  Enter the desired value by using keys 2 (down) and 3 (up). 
Input range:   500...2000 mbar 
5.  Acknowledge the value 
with key 1 (set): The display shows „FilLt“ 
III.) Display damping „FiLt“: 
By this factor the ‘speed’ of the display can be adopted to the application. This is intended to reduce instabilities in the 
measured value as they appear in some specialised applications  
6.  Enter the desired value by using keys 2 (down) and 3 (up). 
Input range:   0...15 (0: fast display, 15: slow display. Setting ex works: 10) 
7.  Acknowledge the value with key 1 (set). 
8.  The settings are stored. The unit starts up again. (8888 in display) 
Attention: 
Bring back the jumper j1 to the ‚parking position‘ after configuration (one contact connected, the 
second contact ‚on air‘)!  
Then the device can be calibrated by key 1 and the configuration data is protected. 
Error and System Messages 
Display 
Description 
Possible fault cause 
Remedy 
FE  1 
measuring range exceeded 
Calibration is wrong 
Wrong signal 
Recalibrate the transducer. 
 
FE  2
 
Measuring values below 
measuring range 
Wrong signal 
Check the connections, cable and electrode. 
FE  7 
System fault 
Error in device 
Disconnect from supply and reconnect. If error 
remains: return to manufacturer 
FE  9 
Input signal is not valid 
Sensor not connected or 
cable defective 
Check sensor, cable and connections 
8.8.8.8 
Segment test 
The transducer performs a display test for 2 seconds after power up. After 
that it will change to the display of the measuring. 
Input signal is not valid 
Permissible input range is 
exceeded 
Check if not a wrong electrode is connected. 
Replace electrode. 
 
Appendix A: Offset settings for different applications 
The Offset setting is supposed to correct measuring deviations caused by different flow speed (v) around the electrode: 
 
For activated sludge basins and at flow speeds > 1m/s a offset setting 
of  ca 
–0,6 should be entered. 
 
In between 0,3 and 1m/s the offset can be read from the diagram. 
0,3m/s correspond to the speed that is achieved by stirring by hand. 
 
Flow speeds below 0,3 m/s are not sufficient to get a correct and 
reproducible  measuring. 
 
 
-0,6
-0,4
-0,2
0
0,3
0,5
0,7
0,9
1,1
1,3
1,5
1,7
1,9
v [m/s]
O
ffs
e
[m
g
/l
]