d) Data logger, recording of measuring values
The recording of the measuring values is started with the REC button. Every 5s (adju-
stable, recording interval), for example, the measuring values are recorded (T1, T2,
T3 and T4) with time registration. The values are filed at a memory location. You can
see the values in the data logger window of the computer.
Proceed as follows to set the recording interval:
Actuate the button "T1 - T2" at the same time when you turn-on the thermometer
and you will get to the setup "Set". Now actuate the button "MAX MIN" once. After
that, there will appear "Int" for Interval and a flashing minute display. Now set the
desired recording interval in minutes and seconds. The maximum recording interval
is 59 minutes and 59 seconds. The minimum value is limited to "00: 01" (= 1s).
Actuate the "MAX MIN" button once again after completed setting and you will get
back to the current temperature measuring display. 
Note!
Each press of a button is confirmed with a short "Beep" of the installed beeper.
C3 Jack assignment
The jacks T1 to T4 are so-called unipolar blade contact jacks. You must connect the
K type temperature sensors to these jacks if you want to carry out temperature mea-
surements according to the sensor specifications. Take into account that the blade
contact jacks have different widths. 
Never try to squeeze the connectors the wrong way (+ and –
mixed up) with force into the jacks. The jacks would be destroyed
definitely and should be replaced. 
The jack "OUTPUT" is the serial RS-232 interface in 3,5 mm stereo jack format. The
assignment can be defined as follows (connector plan):
At the back of the connector there is the ground = GND = reference mass (= refe-
rence potential)
In the centre of the connector there is the contact RX = 5 V high input (= data input)
At the front at the point there is the contact TX = 5 V high output (= data output)
And finally there is the jack DC 9V. You can connect here a power supply unit with
following output data: 9 V direct voltage, if possible, stabilised, with an output cur-
rent of at least 100 mA, a connector outside diameter of 3,5 mm and a connector
inside diameter of 1,35 mm. The polarity: minus "-" inside, plus "+" outside.
D Working position
Always operate the digital thermometer 309 so that you can read the liquid-crystal
display (LCD) or that the digital display points upwards.