Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Hoja De Datos

Los códigos de productos
GDS-2074A
Descargar
Página de 220
 
 
GDS-2000A Programming Manual 
104
 
Example 
:MEASUrement:MEAS3:SOUrce1 CH1 
:MEASUrement:MEAS3:TYPe PK2PK 
:MEASUrement:MEAS3:STATE ON 
:MEASUrement:STATIstics:MODe ON 
:MEASUrement:MEAS3:MINImum? 
1.600E-02 
Returns the minimum measurement result for 
measurement number 3. 
:MEASUrement:MEAS<X>:STDdev
 
Query
 
 
Description 
Returns the standard deviation for the selected 
measurement from the last time the statistics were 
reset. This is a statistics related command. 
Syntax 
:MEASUrement:MEAS<X>:STDdev? 
Related 
Commands 
:MEASUrement:STATIstics:MODe 
Parameter 
MEAS<X>  The automatic measurement number 
from 1 to 8. 
Example 
:MEASUrement:MEAS3:SOUrce1 CH1 
:MEASUrement:MEAS3:TYPe PK2PK 
:MEASUrement:MEAS3:STATE ON 
:MEASUrement:STATIstics:MODe ON 
:MEASUrement:MEAS3:STDdev? 
1.530E-03 
Returns the standard deviation for measurement 
number 3. 
:MEASUrement:STATIstics:MODe 
 
Set
Query
 
 
Description 
Puts the statics measurement results on the display 
or queries whether the statics are displayed.