Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Hoja De Datos
Los códigos de productos
GDS-2074A
COMMAND DETAILS
107
Parameter
<X>
Reference waveform 1, 2, 3 ,4.
<NRf>
Horizontal co-ordinates
Return parameter <NR3> Returns the reference waveform position
Example
:REF1:TIMebase:POSition -5.000E-5
Selects reference 1, and then sets the horizontal
position to -50us.
Selects reference 1, and then sets the horizontal
position to -50us.
:REF<X>:TIMebase:SCALe
Set
Query
Description
Sets or returns the selected reference waveform
time base scale.
time base scale.
Syntax
:REF<X>:TIMebase:SCALe { <NRf> | ?}
Related
commands
commands
:REF<X>:DISPlay
Parameter
<X>
Reference waveform 1, 2, 3 ,4.
<NRf>
Horizontal scale
Return parameter <NR3> Returns the reference waveform scale.
Example
:REF1:TIMebase:SCALe 5.00E-4
Selects reference 1, and then sets the horizontal
scale to 500us/div.
Selects reference 1, and then sets the horizontal
scale to 500us/div.
:REF<X>:OFFSet
Set
Query
Description
Sets or returns the selected reference waveform
vertical position (offset).
vertical position (offset).
Syntax
:REF<X>:OFFSet { <NRf> | ?}
Related
commands
commands
:REF<X>:DISPlay
Parameter
<X>
Reference waveform 1, 2, 3 ,4.
<NRf>
Vertical offset
Return parameter <NR3>
Returns the reference waveform vertical
position.
position.