Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Hoja De Datos

Los códigos de productos
GDS-2074A
Descargar
Página de 220
 
 
GDS-2000A Programming Manual 
148
 
:RECAll:WAVEform
 
Set
 
 
Description 
Recalls a waveform from wave1~wave20 or from 
file to REF1~4.  
Note 
Only *.LSF files can be recalled using this 
command. *.CSV files cannot be recalled. 
Syntax 
:RECAll:WAVEform{W<n> | <file path> 
(“Disk:/xxx.LSF”,”USB:/xxx.LSF”)},REF<X> 
Parameter 
n  
1~20 (Wave1~wave20) 
 
xxx.LSF 
Filename in file path. 
 
<X> 
1,2,3,4 (REF1, REF2, REF3, REF4) 
Example 
:RECAll:WAVEform W1, REF1
 
Recalls the waveform stored in Wave1 to reference 
1.  
:SAVe:IMAGe 
 
Set
 
 
Description 
Saves a screen image to the assigned file path with 
a specified filename. 
Syntax 
:SAVe:IMAGe {<file path> (“Disk:/xxx.PNG”, 
“USB:/xxx.BMP)} 
Related 
commands 
:SAVe:IMAGe:FILEFormat 
:SAVe:IMAGe:INKSaver 
Parameter 
xxx.PNG or 
xxx.BMP 
File name (8 characters max) 
Example 
:SAVe:IMAGe “Disk:/pic1.PNG” 
Saves a screen image named pic1.png to the root 
directory (Disk:/) of the scope. 
:SAVe:IMAGe “USB:/pic1.BMP” 
Saves a screen image named pic1.bmp to the root 