Gw Instek AFG-2025 Function Generator, Frequency Generator AFG-2025 Hoja De Datos

Los códigos de productos
AFG-2025
Descargar
Página de 145
 
 
AFG-2000 Series User Manual 
72
 
 
2.  The Trigger source will be displayed at the 
bottom of the screen.  
 
 
Range  
Source 
INT, EXT 
Connection  
(EXT source only) 
For external sources, 
connect the Sweep trigger 
signal to the Trigger input 
port on the rear panel. 
OUTPUT
INPUT
MOD
Counter
Trigger
MOD
 
Example: 
Source = EXT 
 
Note 
With an external source, a sweep is output each 
time a trigger pulse (TTL) is received from the 
Trigger input port on the rear panel.    
The trigger frequency must be greater than the 
sweep rate (sweep time) plus 125nS (trigger pulse 
width > 125nS).  
Example: External 
trigger input 
signal 
Sweep output
0V
Trigger input 
signal