Rigol DM3058E Digital-Multimeter, DMM, DM3058E Hoja De Datos
Los códigos de productos
DM3058E
3
DC Current
Shunt Resistor
Shunt Resistor
200 μA sampling voltage < 8 mV
2 mA sampling voltage < 80 mV
1Ω for 20 mA, 200 mA
0.01 Ω for 2 A, 10 A
0.01 Ω for 2 A, 10 A
Input Protection
Rear panel: accessible 10 A, 250 V fast-melt fuse
Internal: 12 A, 250 V slow-melt fuse
Continuity / Diode Test
Measurement Method
Measurement Method
1 mA ± 5% constant-current source, < 8 V open-circuit voltage
Response Time
123 samples/sec, with beeper
Continuity Threshold
Adjustable between 1 Ω and 2000 Ω
Input Protection
1000 V
Setup time Attentions
The setup time about voltage measurement is
The setup time about voltage measurement is
influenced by source resistance and media characteristics of cable as
well as input signal
AC Characteristics
Accuracy ± (% of reading + % of range)
[1]
Function
Range
[2]
Frequency Range
1 Year
23℃±5℃
Temperature
Coefficient
0℃ - 18℃
28℃ - 50℃
28℃ - 50℃
True RMS
AC Voltage
[3]
200.000 mV
20 Hz – 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz – 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
3.0 + 0.05
0.05 + 0.010
2.00000 V
20 Hz – 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz – 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
3.0 + 0.05
0.05 + 0.010
20.0000 V
20 Hz – 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz – 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
3.0 + 0.05
0.05 + 0.010
200.000 V
20 Hz – 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz – 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
3.0 + 0.05
0.05 + 0.010
750.000 V
20 Hz – 45 Hz
1.5 + 0.10
0.01 + 0.005
45 Hz – 20 kHz
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
1.0 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
3.0 + 0.05
0.05 + 0.010
True RMS
AC Current
[5]
20.0000 mA
20Hz – 45 Hz
1.5 + 0.10
0.015 + 0.015
45 Hz - 2 kHz
0.50 + 0.10
0.015 + 0.006
2 kHz -10 kHz
2.50 + 0.20
0.015 + 0.006
200.000 mA
20 Hz - 45 Hz
1.50 + 0.10
0.015 + 0.005
45 Hz – 2 kHz
0.30 + 0.10
0.015 + 0.005
2 kHz - 10 kHz
2.50 + 0.20
0.015 + 0.005
2.00000 A
20 Hz – 45 Hz
1.50 + 0.20
0.015 + 0.005
45 Hz - 2 kHz
0.50 + 0.20
0.015 + 0.005
2 kHz – 10 kHz
2.50 + 0.20
0.015 + 0.005
10.0000 A
[5]
20 Hz – 45 Hz
1.50 + 0.15
0.015 + 0.005
45 Hz - 2 kHz
0.50 + 0.15
0.015 + 0.005
2 kHz – 5 kHz
2.50 + 0.20
0.015 + 0.005
Additional wave crest factor error (not Sine)
[6]
Wave crest coefficient
Error (% range)
1 - 2
0.05
2 - 3
0.2
Remarks:
[1] Specifications are for 0.5 hour warm-up, “Slow” measure and calibration temperature 18℃ - 28℃.
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.
[2] 20% over range on all ranges except for DCV 1000 V, ACV 750 V, DCI 10 A and ACI 10 A.