6-7 
Example 3: dBm Measurement 
 
Purpose: 
Execute the dBm measurement through command. 
dBm is frequently used in Audio signal measurement and analysis.   
 
The programming implementations——RIGOL Commands 
0    *RST   
 
 
 
 
/* Reset the instrument */ 
1    cmdset rigol               /* Select Agilent commands */ 
2    *IDN?   
 
 
 
 
/*  Query the equipment ID and confirm if the 
multimeter works normally */ 
3     /* Return RIGOL Technologies,DM3058,DM3A020080808,99.00.00.00.00.00*/ 
4    :function:resistance   
 
/* Enable resistance measuring function */ 
5    :measure AUTO            /* Set the measuring range to “AUTO” */ 
6    :measure:resistance?  
    /* Query the measured resistance */ 
7    /* The return value is used as the reference resistance for dBm measurement 
*/ 
8    :function:voltage:AC   
    /* Enable AC voltage measuring function */ 
9    :measure AUTO   
 
/* Set the measuring range to “AUTO”*/ 
10   :calculate:function DBM          /* Enable dBm measuring function */ 
11   :calculate:DBM:reference <value> /* Set the reference voltage */ 
12   :calculate:DBM?  
 
 
/* Query the measured result / 
13      /* Return value */ 
 
Explanations: 
1.  The operation in line 6 must be done before the next. 
2.  Before sending the command in line 6, the instrument should be connected under 
resistance measure mode. 