Rigol DM3058E Digital-Multimeter, DMM, DM3058E Hoja De Datos
Los códigos de productos
DM3058E
RIGOL
Programming Guide for DM3058
4-1
Chapter 4 Compatible Agilent Commands
In ths commands system, all the commands and the return format are ASCII
characters. Before using them, send the command CMDSET - CMDSET AGILENT
firstly to enable the compatible Agilent commands. The system is case insensitive, and
supports the key words abbreviation (the
abbreviated letters are capitalized to separat
from the whole key words). The compatible commands must follow the syntax strictly,
for instance, unwanted characters are not allowed in command trees; command trees
and parameters should be separated by a space, unwanted characters are also not
allowed to follow parameters, or else the instrument will beep to alarm for an
execution error.
for instance, unwanted characters are not allowed in command trees; command trees
and parameters should be separated by a space, unwanted characters are also not
allowed to follow parameters, or else the instrument will beep to alarm for an
execution error.
The system mainly includes the following sub commands: