Basic Measurements 
24 
 
3.2.2    Crest factor 
AC voltage and current accuracies are affected by the crest factor of the waveform, the ratio of the peak 
value to the RMS value. Table 3-1 lists the fundamental frequencies at which the corresponding crest 
factor must be taken into account for accuracy calculations. 
 
Table 3-1 
Crest Factor Limitations 
Crest Factor 
Fundamental Frequency 
4-5 
50 kHz 
3 kHz 
1 kHz 
3.3    Measuring Current   
Current ranges: 12 mA, 120 mA (DCI only, not available for ACI), 1.2 A, 12 A 
Maximum resolution: 100 nA (on 12 mA range) 
 
 
 
 
 
    
Note:    Auto range is only available for 12 mA and 120 mA (DCI only) ranges.    For 1.2 A and 12 A 
range, manual range must be used. 
 
3.3.1    Connections 
Assuming the multimeter is under factory default conditions, the basic procedure for measuring current 
is as follows: 
1.  Connect test leads to INPUT LO and SENSE LO terminals 
2.  Select DCI or ACI measurement function by pressing 
  →  
  or 
  →  
 
respectively. 
3.  Press 
  to toggle between auto and manual ranging. Notice the AUTO annunciator is displayed 
with auto ranging. For manual range, use the RANGE    
  and       keys to select a 
measurement range consistent with expected current. 
Auto range is only available for 12 mA and 120 mA (DCI only) ranges.    Manual range must be 
used for 1.2 A and 12 A ranges. 
Therefore, it is recommended to use manual range when measuring current greater than 1 A with   
1.2 A or 12 A range. 
4.  Connect test leads to the source as shown in Figure 3-2: 
 
CAUTION:   Do not apply more than 2  A  between  INPUT LO and SENSE LO terminals  or the 
protective  fuse  on the front panel will blow.    Use the 10 A terminal for measuring 
current above 1 A.    See Figure 3-2 for details. 
 
5.  If the “OVR.FLW” message is displayed, press up         key to select a higher range until a normal 
reading is displayed (or press 
  key for auto ranging). Use the lowest possible range for the 
best resolution.    Note that auto ranging is only available for 12 mA and 120 mA (DCI only) ranges.