Remote Operation 
56 
 
5.2    USB & RS-232 Interface Operation 
5.2.1    RS-232 Connection 
The RS-232 interface on this instrument uses a 9-pin DB9 connector.    The pin outs are defined below in 
Table 5-1: 
 
Table 5-1 
RS-232 Pin Out 
Function 
Code 
9 Pin Connector Pin Number 
Transmitted Data 
TXD 
Received Data 
RXD 
Signal Ground Common 
GND 
 
Figure 5-1 shows the rear panel connector for the RS232 interface. 
 
Figure 5-1 
Rear Panel RS232 Interface 
 
Note:    To interface with a serial port on a computer with the RS232 interface, use a NULL 
MODEM or CROSS OVER serial DB-9 female to female cable.    Do NOT use a straight-through 
serial DB-9 cable.    To check that you have the correct cable, probe pin 2 on one end and pin 3 on 
the other and vice versa to check continuity.     
 
Below is a pin connection diagram between 5492B and a computer (Figure 5-2): 
 
                     
 
Figure 5-2 RS-232 Connection Sketch 
 
 
Note: 
Pin 4 and 6, pin 7 and 8 are shorted respectively at the end of the controller.