General 
Capacitors 
Capacitors are usually measured in series except 
for extremely small capacitance. Like with 
inductors, larger capacitors should be measured 
with low frequencies. Small capacitors with high 
frequencies.     
Test 
Frequency 
Expected Capacitance 
<10pF 10pF~400pF  400pF~1uF >1uF 
0.1~0.12 kHz 
 
 
 
Series 
1kHz 
 
 
Series 
 
10kHz 
 
Series or 
Parallel 
 
 
100kHz Parallel 
 
 
 
General Resistors 
A series inductance circuit is the best equivalent 
circuit for low resistance (<1kΩ) and a parallel 
capacitance circuit for high resistances (>10MΩ).  
Test 
Frequency 
Expected Resistance 
<1kΩ 1kΩ~10MΩ >10MΩ 
 
0.03kHz 
 
 
Parallel 
 
0.25kHz 
 
Parallel 
 
 
1kHz Series 
 
 
 
Metal component 
case connection 
A large area of metal can add noise to the 
measurement. Here is how to minimize the effect. 
If the metal is connected to one of the terminals, 
this should be connected to the Hforce terminal 
side. 
If the metal is NOT connected to either of the 
terminals, connect to the GND terminal. 
Wire capacitance 
When measuring the wire capacitance, the fixture 
clips that are marked with H
F
(High Force)/H
S
 
(High Sense)
 
should always be connected to the  
point that is influenced the most by noise.  
   LCR-800 
User 
Manual 
28
 
Air-cored coils 
Air-cored coils can pick up noise very easily, 
therefore they should be kept well clear of any test 
equipment that may contain power transformers or 
display scan circuitry. Also, keep the coils away 
from metal objects which may modify inductor 
characteristics.  
Iron-cored and 
ferrite inductor 
The effective value of iron-cored and ferrite 
inductors can vary widely with magnetization and 
test signal level. Measure them at the AC level and 
frequency in use. Unlike most inductors, a parallel 
equivalent circuit is most suitable for iron-cored 
inductors. When core materials are damaged by 
excessive magnetization (for example: tape heads 
and microphone transformers), check that the test 
signal is acceptable before connection.