  Test Result Format 
Ex: When LCR Meter have Processed Measurement, then will sequent send 2 command for test result automatically. Refer to below: 
Ex: When C=1nF,D= .0045 (C/D mode and display mode is VALUE) 
1. Primary factor of test result. (It doesn’t include primary's unit).   
     LCR Meter response MAIN:PRIM  1.0000^END 
          ASCII CODE    = 4D 41 49 4E 3A 50 52 49 4D 20 20 31 2E 30 30 30 30 0A    (Hex format)     
                                                                                                  Positive symbol 
    2. Secondary factor and primary’s unit 
     LCR Meter response MAIN:SECO  .0045nF^END 
          ASCII CODE    = 4D 41 49 4E 3A 53 45 43 4F 20 20 2E 30 30 34 35 6E 46 0A  (Hex format) 
                                                                                          Positive symbol          Primary factor’ unit   
 
Ex: When R=1ohm,Q= .0005 (R/Q mode and display mode is VALUE) 
  1. Primary factor of test result (It doesn’t include primary's unit.   
     LCR Meter response MAIN:PRIM  1.0000^END 
          ASCII CODE    = 4D 41 49 4E 3A 50 52 49 4D 20 20 31 2E 30 30 30 30 0A    (Hex format)     
                                                                                                  Positive symbol 
    2. Secondary factor and unit 
     LCR Meter response MAIN:SECO  .0005  ^END 
          ASCII CODE    = 4D 41 49 4E 3A 53 45 43 4F 20 20 2E 30 30 30 35 20 20 0A  (Hex format) 
                                                                                          Positive symbol      Primary factor’s unit