INTERACE 
115 
 Pin11 
/O_BIN_11 
Go, Assigned BIN 11 
 Pin12 
/O_BIN_12 
Go, Assigned BIN 12 
 Pin13 
/O_BIN_13 
Go, Assigned BIN 13 
 Pin14 
/O_S_OVER 
No-Go/D or Q fail 
 Pin15 
/O_P_OVER RLC 
FAIL(O) 
 Pin16 
GND GROUND 
 Pin17 
VCC VCC 
 Pin18 
GND GROUND 
 Pin19 
/O_P_HI RLC 
FAIL(O) 
 Pin20 
/O_P_LO RLC 
FAIL(O) 
 Pin21 
/O_S_REJ 
No-Go/D or Q fail 
 Pin22 
/O_INDEX Data 
acquisition 
over, 
OK to remove DUT(O) 
 Pin23 
/O_EOM 
End of Test(O) 
 Pin24 
/I_E_TRIG Start 
Measurement(I) 
 Pin25 
/I_K_LOCK Panel 
Lock 
 
 
   LCR-800 
User 
Manual 
116
 
Signal Overview 
 
Background 
The signal overview section describes the 
functions and the overall characteristics of the 
signals used in the handler interface.  
Parameter Output 
Signals 
 /O_INDEX 
The Index signal will become low 
when the Analog measurement time 
has completed. When the Index 
signal is low, the test component can 
be replaced with the next 
component. The signal goes high 
when the next trigger is active. 
 /O_BIN_1 
/O_BIN_13 
The Bin Go/No-Go signals go active 
low when a successful comparison 
has been made. For example if a 
component is assigned to Bin_1, 
/O_BIN_1 signal goes low until time 
T4. All the remaining signals 
(/O_BIN_2~/O_BIN_13) remain 
high.   
 /O_P_HI 
When the primary measurement is 
higher than the MAX limit, O_P_HI 
will go low until time T4. 
 /O_P_LO 
When the primary measurement is 
lower than the MIN limit, O_P_LO 
will go low until time T4. 
 /O_P_OVER 
When the primary measurement is 
higher or lower than the MAX/MIN, 
O_P_OVER will go low until time T4.