Metrel MD 9040 Digital-Multimeter, DMM, 20991440 Hoja De Datos

Los códigos de productos
20991440
Descargar
Página de 46
MD 9040 & MD 9050 Digital multimeter 
Specification 
 
19 
6 Specification 
 
General Specifications 
 
Display:  
9999 counts: ACV, DCV, Hz & nS 
 
6000 counts: mV, 
A, mA, A, Ohm & Capacitance 
 
Polarity:
  
Automatic 
 
Update Rate:  
Digital Display: 5 per second nominal; 
 
41 Segments Bar-graph: 60 per second max 
 
Low Battery:
  
Below approx. 7V 
 
Operating Temperature:
  
0
o
C to 45
o
 
Relative Humidity:
  
Maximum relative humidity 80% for temperature up to 31
o
decreasing linearly to 50% relative humidity at 45
o
 
Pollution degree:
  
 
Storage Temperature:
  
-20
o
C to 60
o
C, < 80% R.H. (with battery removed) 
 
Altitude: 
Operating below 2000m 
 
Temperature Coefficient:  
nominal 0.15 x (specified accuracy)/ 
o
C @(0
o
C ~ 18
o
C or 
28
o
C ~ 45
o
C), or otherwise specified 
 
Sensing:  
MD 9050: AC+DC True RMS 
 
MD 9040: AC True RMS 
 
Safety:
  
 
Double insulation per IEC61010-1 2nd Ed., EN61010-1 2nd 
Ed., UL61010-1 2nd Ed. & CAN/CSA C22.2 No. 61010.1-
0.92 to Category IV 1000Vac & Vdc. 
 
Transient protection:  
12kV (1.2/50
s surge) 
 
MD 9040 & MD 9050 Terminals (to COM) Measurement Category: 
V : 
 
Category IV 1000Vac & Vdc 
mA
A : 
 
Category IV 600Vac and 300Vdc 
A : 
 
Category IV 600Vac and 500Vdc 
 
Overload Protections:  
MD 9050, MD 9040: 
A & mA: 
   
1A/600Vac, IR 100kA, F fuse 
A: 
   
10A/600Vac, IR 100kA, F fuse 
V:  
 
1050Vrms, 
1450Vpeak 
mV, 
, & Others: 
 
 
600Vdc and Vac rms 
 
E.M.C. : 
Meets EN61326-1:2006 (EN55022, EN61000-3-2, EN61000-
3-3, EN61000-4-2, EN61000-4-3, EN61000-4-4, , EN61000-
4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11) 
In an RF field of 3V/m: 
Capacitance function is not specified