MD 9040 & MD 9050 Digital multimeter 
Operation 
 
*Ghost-voltage Buster: 
Ghost-voltages are unwanted stray signals coupled from 
adjacent hard signals, which confuse common multimeter voltage measurements. Our 
AutoCheck
TM
 mode provides low (ramp-up) input impedance (approx. 3kΩ at low 
voltage) to drain ghost voltages leaving mainly hard signal values on meter readings. It 
is an invaluable feature for precise indication of hard signals, such as distinguishing 
between hot and open wires (to ground) in electrical installation applications. 
 
WARNING: 
AutoCheck
TM
 mode input impedance increases abruptly from initial 3kΩ to a few 
hundred kΩ’s on high voltage hard signals. “LoZ” displays on the LCD to remind the 
users of being in such low impedance mode. Peak initial load current, while probing 
1000VAC for example, can be up to 471mA (1000V x 1.414 / 3kΩ), decreasing abruptly 
to approx. 3.1mA (1000V x 1.414 / 460kΩ) within a fraction of a second.  Do not use 
AutoCheck
TM
 mode on circuits that could be damaged by such low input impedance. 
Instead, use rotary selector   or 
 high input impedance voltage modes to minimize 
loading for such circuits.  
 
dBm 
+Hz
 (for MD9050 only), Hz 
+ACV
, ACV 
+Hz
  functions 
Press the SELECT button momentarily to select the subject functions in sequence. Last 
selection will be saved as power up default for repeat measurement convenience. 
 
 
 
Note: Hz Input sensitivity varies automatically with voltage (current function alike) range 
selected. 1V range has the highest and the 1000V range has the lowest. Auto-ranging 
measurements normally set the most appropriate trigger level. You can also press the 
RANGE button momentarily to select another trigger level (voltage range) manually. If 
the Hz reading becomes unstable, select higher voltage range to avoid electrical noise. 
If the reading shows zero, select lower voltage range. 
 
Note: In dBm 
+Hz
 function, power up default reference impedance will be displayed for 1 
second before displaying the dBm readings. Press dBm-
 (RANGE) button momentary 
to select different reference impedance of 4, 8, 16, 32, 50, 75, 93, 110, 125, 135, 150, 
200, 250, 300, 500, 600, 800, 900, 1000, up to 1200
. Last selection will be saved as 
power up default for repeat measurement convenience. Manual trigger level selection 
on Hz reading is not available.