Gw Instek GDM-8342USB Digital-Multimeter, DMM, GDM-8342USB Hoja De Datos

Los códigos de productos
GDM-8342USB
Descargar
Página de 158
 
 
GDM-834X Series User Manual 
128
 
[SENSe:]FREQuency:INPutjack 
Assigns an input terminal for the frequency function. 
Parameter: (0|1|2) 0=volt, 1=500mA, 2=10A 
Example: SENS:FREQ:INP 0 
Sets the input jack to the Volt input terminal. 
[SENSe:]FREQuency:INPutjack? 
Returns the assigned input terminal used for the frequency 
function. 
Return Parameter: VOLT, 500mA, 10A 
[SENSe:]PERiod:INPutjack 
Assigns an input terminal for the period function. 
Parameter: (0|1|2) 0=volt, 1=500mA, 2=10A 
Example: SENS:PER:INP 0 
Sets the input jack to the Volt input terminal. 
[SENSe:]PERiod:INPutjack? 
Returns the assigned input terminal used for the period function. 
Return Parameter: VOLT, 500mA, 10A 
[SENSe:]CONTinuity:THReshold 
Sets the continuity threshold in ohms. 
Parameter: <NRf> (0 ~ 1000) 
Example: SENS:CONT:THR 500 
Sets the continuity threshold to 500 
[SENSe:]CONTinuity:THReshold? 
Returns the continuity threshold.