Table 6-1 
DISPlay Subsystem Commands Summary 
Command 
Function Description 
:DISPlay 
:ENABle <b> 
  :ENABle? 
 
Enable or disable front panel display 
Query state of the display 
 
:ENABle <b> 
Command syntax: 
:DISPlay:ENABle <b> 
 
Command Parameter: 
<b> = 
0 or OFF 
 
Disable front panel display 
 
 
    1 or ON   
 
Enable front panel display 
Query: 
:ENABle?                Query state of the display 
 
Description: 
This command is used to enable or disable front panel display circuitry. When disabled, the 
instrument operates at a higher speed. While disabled, the display is frozen. All front panel controls 
except LOCAL are disabled. Normal display operation can be resumed by using:ENABle command 
or pressing LOCAL key to enable the display.