:RANGe <n> 
Command syntax: 
:FREQuency:THReshold:VOLTage:RANGe <n>    Set voltage threshold range. 
:PERiod:THReshold:VOLTage:RANGe <n>   
Set voltage threshold range. 
 
Command parameter: 
<n> =    0 to 1010 
 
 
Specify signal levels in volts (voltage threshold) 
 
Query: 
:RANGe? 
 
 
 
 
Query maximum signal level. 
 
Description:   
These commands are used to specify the expected input level. The  instrument will then 
automatically select the most ideal current or voltage threshold range. 
 
:REFerence <n> Commands 
:REFerence <n> 
Command syntax: 
:FREQuency:REFerence <n>   
 
 
Specify reference for FREQ 
:PERiod:REFerence <n>           
 
Specify reference for PER 
 
Command parameter: 
<n> =   
0 to 1.0e6               Reference for FREQ 
             0 to 1                    Reference for PER 
DEFault               
0 (All measurement functions) 
          
MINimum                Minimum value for specified function 
MAXimum               Maximum value for specified function 
Query: 
:REFerence?                  
Query reference for relative function 
 
Description: 
These commands are used to establish a reference value for the specified function. When reference 
 
is enabled (:REFerence:STATe), the result will be the algebraic difference between the input signal 
 
and the reference value: 
 
Reading = Input signal – Reference 
 
From the front panel, reference is called relative (REL). 
 
The  :REFerence <n> command is coupled to the :ACQuire command. The last command sent 
(:REFerence <n> or :ACQuire) establishes  the reference. When a reference is set using 
the :REFerence <n> command, the REFerence? query command returns the programmed value. 
Conversely, when a reference is set using the  :ACQuire  command, the :REFerence?  query 
command returns the acquired reference value.