Gw Instek GDM-8341 Digital-Multimeter, DMM, GDM-8341 Hoja De Datos
Los códigos de productos
GDM-8341
OPERATION
49
Set Continuity Threshold
The continuity threshold defines the maximum resistance allowed
in the DUT when testing the continuity.
in the DUT when testing the continuity.
Range
Threshold
0 to 1000Ω (Default Threshold:10Ω)
Resolution
1Ω
Procedure
1. Press MENU.
2. Go to the MEAS menu on level 1
3. Go to the CONT menu on level 2
4. Set the continuity threshold level.
5. Press the Enter key to confirm the continuity
settings.
6. Press EXIT to exit the CONT setting menu.
Display
W
Continuity
setting
Continuity function
indicator