This section provides supplemental information for 
user consideration when operating the LCR meters.  
Some recommendations and explanations are 
provided to help aid in the use of some functions and 
features, in which can help the user gain optimal and 
accurate measurement results. 
 
Selecting Test Frequency 
Test frequency can greatly affect the results of 
measurement reading, especially when measuring 
inductors and capacitors.  This section provides some 
recommendations and suggestions to consider. 
Capacitance 
When measuring capacitance selecting, the right 
frequency is important in obtaining the most accurate 
measurement results.  Generally, a 1 kHz test 
frequency is used to measure capacitors that are 0.01 
µF or smaller.  For capacitors that are 10 µF or larger, 
a lower frequency of 120 Hz is used.  Following this 
trend, high test frequencies are best for testing very 
low capacitance components.  For large capacitance 
components, low frequency would be optimal.  For