Bk Precision BK-879B Digital-Multimeter, DMM, LCD 40000 counts BK-879B Hoja De Datos

Los códigos de productos
BK-879B
Descargar
Página de 104
 
 
85 
 
provide an option for DC resistance measurements.  
Therefore, when measuring a resistive component 
that is designed to be measured with DC, readings will 
be incorrect or inaccurate.  Before using the meter to 
measure resistance, please verify whether the DUT 
(device under test) requires DC or AC resistance 
measurement method.  Depending on the method, 
results will vary greatly. 
 
Guard Terminal 
One of the input sockets and terminals is labeled as 
GUARD”.  This terminal does not have to be used in 
all instances for the meter to make measurements.  
But in some instances, it is very useful.  Guard 
terminal generally serves two purposes.   
 
If user is using test leads, the guard terminal can be 
used to connect to the shielding of the test leads.  
Doing so can be useful when making large resistive 
component measurements.  For example, when 
measuring a 10 M
Ω resistor with test leads, at the 
high range the reading may seem to be unstable as a 
few digits may continuously be changing.  Having the 
shield of the test leads connected to the guard 
terminal will help stabilize the reading in some 
instances.