c) Resistance measurement 
   
Make sure that the circuit components, circuits and components to be measured or any other 
measuring objects are voltage-free and discharged.
Proceed as follows to measure resistance: 
-   Turn the rotary switch to the measuring range Ω. The small current values beneath 
the measuring ranges indicate the measuring current.
-   Connect the black test lead to the COM socket (6) and the red test lead to the VΩ 
socket (7).
-  Zero the multimeter using the adjustment screw (3).
-  Make sure that the “Vrms/Vp-p” button (9) is not pressed.
-   Connect both test leads and wait until the scale indicator has stabilized.  The 
display should show a value of approx. 0 ohm. Zero the meter to 0 ohm using the 
0 ohm adjustment knob (8) in case of deviation. Always zero the multimeter when 
changing measuring ranges. If a zero adjustment cannot be conducted, the mignon 
(AA) batteries are in need of replacement.
-   Read the measured value from the “Ω” scale. Multiply the displayed value by the measuring range to obtain the 
correct reading (e.g. 100Ω (display) x10kΩ (measuring range) = 100 x 10,000 = 1MΩ (measured value). 
-  After measurement, turn the rotary switch to position “OFF” to turn off the multimeter.