Alternating current (only with VC170-1), overload protection 1A/250V + 10 A/250 V, Average recording for sinus signal.
Range (40 - 400 Hz)
Accuracy
Resolution
400 
µA
±(1.6 % + 5)
0,1 
µA
4,000 
µA
µA
40 mA
±(2.0 % + 8)
0.01 mA
400 mA
0.1 mA
4 A
±(2.6 % + 4)
0.001 A
10 A
0.01 A
Resistance, overload protection 250 V, test voltage ca. 0.5 V
Range
VC130-1/150-1
Accuracy
Resolution
Range VC170-1
Accuracy
Resolution
200 
Ω
±(1.0 % + 10)
0.1 
Ω
400 
Ω
±(1.6 % + 3)
0,1 
Ω
2000 
Ω
Ω
4 k
Ω
±(1.3 % + 2)
0.001 k
Ω
20 k
Ω
0.01 k
Ω
40 k
Ω
0.01 k
Ω
200 k
Ω
0.1 k
Ω
400 k
Ω
0.1 k
Ω
20 M
Ω
±(1.3 % + 7)
0.01 M
Ω
4 / 40 M
Ω
±(2.0 % + 8)
0.001 / 0.01 M
Ω
Temperature (only VC150-1)
Range
Accuracy
Resolution
-40 to 0 °C
±(10.4 % + 7)
1 °C
>0 to 400 °C
±(3.3 % + 4)
>400 to 1000 °C
±(3.9 % + 4)
Frequency/duty cycle (only with VC170-1), overload protection 250 V
Range
Accuracy
Resolution
10 Hz - 10 MHz  
max. 10 Vrms
±(0.7% + 4)
0.01 Hz - 0.01 MHz
Sensitivity:
< 100 kHz = 300 mV
> 100 kHz = 600 mV
0.1 – 99.9%
0.1%
Acoustic continuity tester 
 
<10 
Ω Permanent sound
Diode test test voltage 
 
Uo 3.0 V 
Diode overload protection/continuity tester: 
250 V
Transistor test “hFE” 
 
0 – 1000ß, test voltage Uce 3 V, test current Ibo 10 
µA 
NCV voltage test 
 
230 V/AC
PACKAGE CONTENT
Digital multimeter // Measurement lines // Battery // Operating instructfions
This data sheet is published by Conrad Electronic SE, Klaus-Conrad-Str. 1, D-92240 Hirschau (www.conrad.com).
All rights including translation reserved. Reproduction by any method, e.g. photocopy, microfilming, or the capture in electronic data processing systems 
require the prior written approval by the editor. Reprinting, also in part, is prohibited. This data sheet represents the technical status at the time of printing.
© Copyright 2014 by Conrad Electronic SE.
V2_0914_02-ETS-Mkde